Tests and measurement entry
A test holds the protocol and the measured values of one exercise test (multi-stage test) of a subject. This chapter describes how to enter a new test, how to enter the values during the test (live input) or import them from a Lactate Scout, and how to open, compare and delete tests.
Starting a new test
Choose (Ctrl+M), click the symbol in the toolbar or New test in the quick help. Ergonizer first asks for the subject in the dialog Select subject, even if a subject is already loaded.
You can also start a test without the selection list:
in the subject dialog with New test results directly >>, see Subjects and groups;
with for the subject whose test is shown.
If no subject has been created yet, Ergonizer asks you to create one first.
The Measurement data dialog
For a new test, the dialog is titled New test; for an existing test, Edit measurement data. It has four tabs, which you normally fill in from left to right:
Basic data – body weight, height and further body data
Exercise protocol – type of ergometry and protocol
Further information – options for this test and for the printout
Measured values – the measured values of each stage
Use the arrows on the left and right or click a tab to change between them. Ergonizer checks the entries of each tab before moving on. You can only move beyond Exercise protocol once a protocol has been selected.
Each tab name shows the state of the tab:
a green tick: nothing is missing;
a filled mark such as 2 missing: required entries are missing;
an outlined mark such as 1 recommended: recommended entries are missing, for example the test location or the lactate analyzer.
The tooltip of the tab names the fields (Still missing: or Recommended:).
The header of the dialog shows the subject (read-only), the Test No. and the Test date. For a new test, the test date is set to today and can be changed.
Cancel closes the dialog without saving.
Basic data
Body weight and Height are required. Ergonizer uses the body weight for relative values such as W/kg (Relative Power), and weight and height for the BMI and the estimated \(\dot{V}\mathrm{O}_{2}\mathrm{max}\).
If the subject already has tests, weight and height of the most recent test are suggested in grey. Change the values or confirm them; if they remain grey when you leave the tab, Ergonizer asks whether they should be taken over unchanged. Answer No to correct them; the cursor is then placed in the field.
The other fields are optional:
Field |
Range |
Used for |
|---|---|---|
Body fat (%) |
2–70 % |
Shown on the printout |
Training mileage |
0–2000 km per week |
Running predictions; without an entry, Ergonizer estimates the training volume. The value is stored in kilometres; Standard unit for distance only changes the display. Values entered in miles in earlier versions are read as kilometres, so enter them again |
Training duration |
0–80 h per week |
Stored and exported |
Vital capacity (L) |
0.5–9 L |
Stored and exported |
Heart volume (mL) |
100–1800 mL |
Stored and exported |
Metric values switches the entry between kilograms/centimetres and pounds/feet/inches. Ergonizer always stores the values in metric units. Whether the option can be changed depends on the settings, see Settings, view and help.
Exercise protocol
Select the protocol in three steps:
Category: laboratory test or field test (Field tests).
Ergometer: for laboratory tests, for example treadmill, cycle ergometer, treadmill walking, handbike, uphill running, rowing ergometer, cross trainer, canoe ergometer, roller ski, Sit-Ski or ski mountaineering test; for field tests, for example running, fencing, swimming, canoe, rowing and roller ski (classic or skating).
Protocol: a template (Competitive sports, Recreational sports, Rehabilitation, Patients), New protocol, an irregular protocol or Lactate vs heart rate.
Tests on the cycle ergometer, handbike, rowing ergometer, cross trainer and canoe ergometer are entered in watts; treadmill tests and all field tests in speed.
Protocol fields
- First stage
The load of the first stage.
- Increment
The increment from stage to stage.
- Stage duration
The step duration, at least 1 minute.
- Maximum stage
The load of the last stage reached. It must be the first stage plus a whole number of increments; a test has 3 to 25 stages.
- Duration max. stage
How long the last stage was sustained; at most the stage duration. If it is shorter than the stage duration, Ergonizer calculates the maximum power proportionally, see Informationen zur metabolischen Leistungsdiagnostik.
From these fields Ergonizer creates the rows of the table on Measured values: one row for the values at rest and one row for each stage.
The templates only fill in first stage, increment and stage duration with typical values, for example on the cycle ergometer:
Template |
First stage |
Increment |
Stage duration |
|---|---|---|---|
Competitive sports |
80 W |
20 W |
3 min |
Recreational sports |
50 W |
50 W |
3 min |
Rehabilitation |
25 W |
25 W |
3 min |
Patients |
25 W |
25 W |
2 min |
You can change all values. The selected template is saved with the test; it has no influence on the evaluation and the training zones. Always enter the maximum stage yourself.
Irregular protocol
Use the irregular protocol for tests with unequal stages. Instead of the protocol fields, a table appears with the load (or distance) and the duration of each stage:
at least 3 stages, each lasting between 15 s and 20 min;
no two stages with the same load; the stages are sorted by load.
Add stage adds a row with 3:00 min (at most 33 stages), Remove stage deletes the selected row, and Repeat value copies the value of the selected cell into all rows of this column.
For field tests with an irregular protocol, enter the distance and time of each stage; Ergonizer calculates the speed.
Lactate vs heart rate
With the corresponding licence module, you can enter the mean heart rate of each stage instead of the load. Ergonizer then evaluates lactate against heart rate instead of power.
Treadmill walking, uphill running and roller ski
For these types you choose what remains constant:
Fixed speed: the speed stays the same (Constant speed), the stages increase the slope.
Fixed slope: the slope stays the same (Constant slope), the stages increase the speed.
For new uphill running tests, the fixed speed and the fixed slope are preset from the settings (8 km/h and 5 %, see Settings, view and help). If you switch between the two, the test keeps both values.
Further information
All options on this tab apply to this test and are saved with it. For new tests, they are preset from the settings (Settings, view and help).
- Lactate constant
The net increase in lactate above the value at the Lactate Threshold that defines the individual anaerobic threshold (see Informationen zur metabolischen Leistungsdiagnostik). Allowed are 0.5 to 5.0 mmol/L; Ergonizer rounds the value to one decimal place. If the field is empty, the default of the settings applies.
- Scheme of training recommendations
The scheme for the training zones, see Training zones and predictions.
- Unit for speed, Unit of measurement for slope
km/h, m/s or mph, and % or degrees. Only shown if the units can be selected individually in the settings. When you change the unit of an existing test, Ergonizer converts the protocol values and warns about rounding errors.
- Test location, Lactate analyzer
Free text; Ergonizer suggests the entries you made before.
- V̇O₂max measured
The maximum oxygen uptake measured by spirometry, in ml/min/kg (5–100).
- Room Temperature
Taken over from the Lactate Scout when importing.
- Treadmill slope
For tests on the treadmill, the ski mountaineering test and the Sit-Ski, unless the slope itself increases from stage to stage; preset from the settings.
- Printout options
Race result predictions, CP curve, estimated \(\dot{V}\mathrm{O}_{2}\mathrm{max}\) and percentiles on the printout. Some of them are only shown if they are enabled in the settings.
- Result prediction w/o lactate
Evaluates the test without lactate, see Evaluation without lactate.
- No automatic correction with difficult LT-estimation
By default, Ergonizer corrects the training recommendations if the lactate curve is atypical. Select this option to suppress the correction.
Measured values
The table contains one row for the values at rest (Rest) and one row for each stage. Depending on the type of ergometry and the licence, it has the following columns:
Column |
Notes |
|---|---|
Stage and load |
Created from the protocol; cannot be edited |
Lactate (mmol/L) |
Hidden for an evaluation without lactate |
Heart rate |
In beats per minute |
Blood pressure |
Entered as systolic/diastolic, for example |
RPE general, RPE muscles, RPE breathing |
RPE on the Borg Scale 6–20 (or 1–10 with Borg CR10 in the settings); not at rest |
Glucose |
Only with the glucose module |
Cadence / Stroke rate |
Only for cycle ergometer, handbike, cross trainer, rowing and canoe; not at rest |
The ? button shows the RPE scale selected in the settings.
Entering values
Ergonizer checks each value as soon as you leave the cell, and all values again when you finish the entry. The messages name the value, the stage and the allowed range, for example:
lactate up to 14 mmol/L at rest and in the first stage, otherwise up to 28 mmol/L;
heart rate 25–200 bpm at rest, higher under load;
glucose 50–500 mg/dL;
blood pressure: the diastolic value must be lower than the systolic value.
Invalid cells are marked in red.
Keyboard:
Key |
Effect |
|---|---|
Enter, Tab |
Confirm the cell and move one row down; in the last row, to the next column |
Shift+Tab |
Move back |
Arrow keys |
Move to the neighbouring cell |
Esc |
Same as Back |
Minimum number of lactate values
A lactate curve requires at least four lactate values, including the value at rest. With fewer values, Go to evaluation stays inactive, and Ergonizer offers an evaluation without lactate. Likewise, at least four heart rate values, including the value at rest, are required.
If the values under load are complete but the value at rest is missing, Ergonizer offers to estimate it.
Evaluation without lactate
With Model based prediction without lactate (on Further information: Result prediction w/o lactate), Ergonizer does not calculate a lactate curve. Instead, a statistical model estimates the thresholds from age, body weight, maximum power, maximum heart rate, slope and sport. The screen and the printout show the note Model based result prediction without lactate measurement!.
Post-exercise values
Expand Edit post-exercise data below the table to enter lactate, heart rate, blood pressure and glucose during recovery. Ergonizer uses them for the recovery curve and the Stegmann threshold. Add stage adds a recovery stage (suggested times 1, 3, 5, 7, 10, 15, 20, 30, 45 and 60 min, which you can change); Remove stage deletes the selected one. At most 10 recovery stages are possible, and each recovery stage with values needs a time.
Lactate outliers
With Exclude outliers from the curve (on by default), Ergonizer checks the open test for lactate values that deviate clearly from the course of the other values. This requires at least five measuring points under load. The first and the last stage and the post-exercise values are not checked automatically, and neither are tests entered as lactate against heart rate. Tests without lactate have no such switch.
If a value deviates, Ergonizer asks Exclude the lactate value of stage … as an outlier?: Exclude leaves it out of the curve; Keep or closing the dialog switches the automatic check off for this test.
Ergonizer can also suggest up to three values together, at most one third of the measuring points and only if at least five values remain: two values from seven measuring points on, three from nine. Exclude and Keep then apply to all the values named; you cannot select individual values in this dialog. The check does not find every outlier: two neighbouring high values, for example, are not always recognized. You can exclude such values yourself.
If a test already has excluded values, Ergonizer suggests no new ones and only checks the excluded values. If the value of an inner stage fits its neighbours, it asks Use the excluded lactate value of stage … again?. Use again includes the value again; if another value deviates clearly, the dialog names it, and that value is excluded instead. Leave unchanged or closing the dialog changes nothing and leaves the check switched on.
If you switch Exclude outliers from the curve off, all excluded values count again, including those you excluded yourself, but the exclusions stay stored. When you switch it on again, they apply again. If no exclusions are stored, Ergonizer then searches again and, if it finds nothing, reports No lactate value was found that can clearly be excluded as an outlier. The setting is saved for each test immediately. The switch is also available in the results panel, see Lactate curve and results.
An excluded value remains stored. It is only left out of the curve and marked Outlier in the chart. You can also exclude or reinstate a value yourself: right-click the lactate point in the chart and choose Exclude from the curve as an outlier or Reinstate outlier – also for the first and the last stage, as long as at least four values remain; with only four active values left, the menu item is greyed out. Of the post-exercise values, at least one must remain. If the switch is off, excluding a value yourself switches it on, and the stored exclusions apply again as well. If excluding a value makes the Stegmann threshold impossible, Ergonizer warns you. If you change an excluded value in the table, it is included again.
See Lactate curve and results for the effect on the evaluation.
Live input during a running test
Live input is meant for entering the values while the test is running, when the last stage is not yet known.
Fill in Basic data and select the protocol with first stage, increment and stage duration on Exercise protocol. A maximum stage is not needed.
Click Live input. The window Enter measured values opens with the row for the values at rest and the first stage. Ergonizer fills in the load of each new stage.
Enter the values of each stage as soon as they are available. Moving past the end of a row with Enter or Tab creates the next stage. The current row is highlighted; completed rows appear in bold.
At the end of the test, click Finish entry and enter the Duration of last stage. Ergonizer takes the highest stage reached as the maximum stage and shows the values on Measured values for checking.
Click Go to evaluation.
Live input has no clock; the stages advance only as you enter the values. Live input is available for new tests only.
Importing lactate values from the Lactate Scout
For new tests, Import Lactate Scout Data reads the lactate values from a Lactate Scout. The button is active once the protocol is complete, including the maximum stage. Ergonizer first asks how the device is connected: via Bluetooth (Lactate Scout 4 or Lactate Scout Sport) or by cable (Lactate Scout+ on the serial port).
Bluetooth
Requirements: Windows 10 version 1809 or later and a Bluetooth LE adapter. The Lactate Scout must be within about 3 m of the computer.
Measure the lactate values in the step test mode of the Lactate Scout: the value at rest in the pre-load phase and one value per stage in the load phase (see the instruction manual of the device). Ergonizer takes over the value at rest and the stage values of the most recent step test, determined by the date and time of the device – so make sure that the clock of the Lactate Scout is set correctly. An older step test cannot be selected. Single and control measurements, invalid measurements and the values of the after-load phase are not taken over; enter post-exercise values by hand.
On the Lactate Scout, switch on Bluetooth: in the menu Device settings, select the Bluetooth symbol and then the activation symbol. The Bluetooth symbol appears in the info line.
Bluetooth address: the device shows it in Device settings under the device information (info symbol); browse with the arrow keys to MAC address of the Bluetooth LE module. You only need to look it up once; note it down, for example on the device.
In Ergonizer:
In the dialog Lactate Scout – Bluetooth Import, enter the Bluetooth Address (hex):: 12 hexadecimal digits, with or without colons or hyphens.
Click Connect & Read. The status line shows date, time and number of lactate values of the step test found. If the device contains no step test with lactate values, Use data remains inactive.
Click Use data.
If the message The data could not be read from the Lactate Scout. appears, check that Bluetooth is switched on at the device, that the device is within about 3 m and that the address is correct, and click Connect & Read again.
Cable
Set the COM port of the Lactate Scout+ in the settings, see Settings, view and help. Ergonizer splits the stored samples into series wherever more than 10 minutes lie between two samples; browse with Next dataset >>. Clear Use value for values you do not want to take over, select Resting lactate included if the first value is the value at rest, and click Use selected data.
After the import
Ergonizer assigns the values to the stages in their order, not by time. If a stage duration is stored in the device, Ergonizer asks whether to use it. The room temperature is taken over as well. If fewer lactate values than stages arrive, Ergonizer warns you. The import does not supply heart rate; enter it and the other values by hand on Measured values.
Evaluating and saving
Go to evaluation on Measured values checks the values, saves the test and shows the evaluation in the main window. The button is only active when the values can be evaluated. Back goes back to the previous tab without saving.
There is no separate save command: a new test is saved with Go to evaluation. When you have changed an existing test, Ergonizer asks Do you want to save the changes? and lists each change with the old and the new value. Choose Save changes or Discard changes. Without changes, there is no question.
Opening, comparing and deleting tests
Open and edit a test
Choose (Ctrl+O) and select the subject. The list only contains subjects with tests and shows the type of ergometry of the last test (● with lactate, ○ without lactate). The dialog Edit measurement data opens with the most recent test; select another test of the subject under Test No..
If a test of the subject cannot be read, Ergonizer shows the note Incompletely saved test hidden with its number and date. The test is hidden, not deleted; the Ergonizer support can restore it.
If you have only changed entries on the first three tabs, for example the body weight, the lactate constant or the training scheme, click Go to evaluation directly on that tab: Ergonizer evaluates the stored values again without showing the table.
opens the test shown, without the selection list.
Compare with an earlier test
The Test selection and results overview at the bottom of the main window lists the tests of the subject. Its columns control the comparison:
1st test: the test shown.
2nd test: the comparison test, shown as a second curve. It must be of the same type of ergometry and have at least 3 stages. Ergonizer preselects the most recent other test of this type, or the test of about one year earlier if Automatically search for previous year’s examination is selected in the settings.
Tests for the summary evaluation: the tests for the summary table on the printout.
The comparison curve is calculated with the lactate constant of the test shown.
For cycling tests evaluated without lactate, \(W'\) in the overview follows the Critical Power value of the report and can show missing.
Delete a test
Open the test with , select it under Test No. and click the waste bin (Delete data set permanently). Ergonizer asks whether to delete the test of this subject, then deletes it with its stage values, training zones, comment, group assignments and calculated results, and confirms The test has been deleted.. The remaining tests of the subject are renumbered.
Warnung
Deleting a test cannot be undone.
Trial version
The trial version stores at most 6 tests of your own; the supplied sample tests do not count. When the limit is reached, New test shows a note instead of the dialog. You can still open, change and delete your own tests.
The sample tests can be opened and evaluated, but not changed or deleted (Sample data cannot be changed or deleted); only the smoothing, the zone boundaries, the comment and the lactate outliers can be saved. A new test for a sample subject is possible and counts as one of your own tests.
Comments and text modules
Comment: Tools ‣ Comment opens a text field for a comment or interpretation of the test shown (one comment per test, up to 900 characters). The comment is printed on page 3 of the report.
Text modules: Tools ‣ Text modules copies a prepared text about the test shown to the clipboard, so that you can paste it into another program, for example a letter, with Ctrl+V:
Performance diagnostics: type of ergometry, protocol, individual anaerobic threshold with heart rate, and maximum power;
Subject data: name, date of birth and place of residence;
Training intensities: the training zones according to the selected scheme.
You cannot create your own text modules.